Minimum load cell verification interval Vmin = EMax / 10,000, full scale output 2 mV/V ±0.1% (1.3 mV/V for CPA150), repeatability 0.01% F.S., combined error 0.025% Emax for capacities ≤10 t and 0.05% Emax for capacities ≥20 t, temperature effect on zero 0.0014%/°C, temperature effect on full scale output 0.0013%/°C, compensated temperature range −10°C/+40°C, operating temperature range −30°C/+60°C with optional extension up to 200°C, creep error after 4 hours 0.03% F.S., maximum tolerated excitation voltage 15 Vdc, input resistance 1100 ±50 Ohm, output resistance 1000 ±10 Ohm, zero balance ±2% F.S., insulation resistance greater than 5000 MOhm, safe overload 150% F.S., and supplied with a 6-wire shielded cable 5 mm in diameter and 10 m long for versions from 150 to 10,000 kg as well as for versions from 20,000 to 50,000 kg.