Minimum load cell verification interval Vmin = EMax / 10,000, maximum number of verification intervals nLC = 3000, full scale output 2 mV/V ±10%, combined error 0.017% of full scale output, temperature effect on zero 0.0014%/°C, temperature effect on full scale output 0.0014%/°C, compensated temperature range −10°C/+40°C, operating temperature range −20°C/+60°C, creep error after 30 minutes 0.025% F.S., maximum tolerated excitation voltage 15 VDC, input resistance 410 ±20 Ohm, output resistance 350 ±5 Ohm, zero balance ±10% F.S., insulation resistance greater than 1000 MOhm, safe overload 150% F.S., ultimate overload 200% F.S., nominal displacement less than 1 mm, and supplied with an SPBC 6-wire shielded cable, 3 m long and 5 mm in diameter.